GUGU-P10
UV-LED Point Light Source
Ø10 mm spot, up to 2,000 mW/cm². Air-cooled head on 24 V DC, ideal for manual dispensing cells.
- 365 / 385 / 395 / 405 nm
- Ø10 mm
- up to 2,000 mW/cm²
Medical device assembly uses UV-LED curing because it bonds in seconds at room temperature, with no heat to distort thin-walled tubing or damage sterilised packaging. Point sources handle catheter and needle joints; small area windows handle sensor lids and connector shells.
Heat is the enemy in medical assembly. Anything that raises the part above roughly 60 °C risks warping a catheter lumen or softening a bond line that was just set. UV curing adds energy without adding heat load to the part, because the cure is photochemical rather than thermal.
The other advantage is speed: a bond that would hold a fixture for 30 minutes with a two-part epoxy is released in seconds, which is what makes high-volume disposable production economically possible.
Only where light can reach the adhesive. For opaque housings, cure the bond before closing the assembly, or use a dual-cure adhesive that also reacts to moisture so shadowed areas finish curing over 24–48 hours.
The light source itself is a tool, not a certified process. What matters for validation is a documented, repeatable dose — which is why we recommend logging irradiance and dwell time per station and re-checking with a radiometer on a fixed schedule.
GUGU-P10
Ø10 mm spot, up to 2,000 mW/cm². Air-cooled head on 24 V DC, ideal for manual dispensing cells.
GUGU-A4020
Small 40 × 20 mm window at high irradiance for sensor and connector bonding.
Send the adhesive datasheet plus part size and cycle time. We reply with emitter type, wavelength and dose.
Ask an engineer